Cleaning is the physical removal of debris. It has two major effects. First, it
reduces the number of micro-organisms present. Second, it removes organic matter,
such as blood, tissue and other debris that may interfere with sterilization or
disinfection. In some instances, cleaning is all that is necessary. Most often,
however, it is the preliminary step before sterilization or disinfection. In these
instances, it is referred to as pre-cleaning. Pre-cleaning is essential, because
sterilization and disinfection procedures may not be effective if items have not
been cleaned first.
Sterilization is the process that destroys all types and forms of micro- organisms
including viruses, bacteria, fungi and bacterial endospores. Major methods of sterilization
include the use of steam under pressure (steam autoclave), dry heat and unsaturated
chemical vapour.
Disinfection is generally less lethal to pathogenic organisms than sterilization.
Three levels of disinfection have been differentiated, depending on the type and
form of micro-organisms destroyed.
The effectiveness of any disinfection procedure is influenced by several factors,
including the type and number of micro-organisms present, the concentration and
length of exposure to the disinfecting agent and the amount of organic matter or
other debris present on the item being disinfected. For disinfecting agents to be
effective, it is important that they are used according to the manufacturer's instructions.
When to Sterilize, Clean or Disinfect patient care items
The major factor determining whether an item must be sterilized, disinfected or
simply cleaned is how it is used.
Items To Be Sterilized or Disinfected :-
- Instruments that penetrate oral soft tissue (the mucosa or skin) or bone must be
sterilized after each use or discarded.
- These items are termed critical items
and have the greatest risk of transmitting infection, for e.g. surgical instruments,
periodontal knives and scaling instruments.
- Instruments that are not intended to penetrate oral soft tissues or bone (such as
amalgam condensers, dental handpieces, mouth mirrors) but come in contact with mucous
membranes or non-intact skin are termed semi critical items.
- These items pose a lower risk of disease transmission; however, because the majority
of semicritical items in dentistry are heat tolerant, they also should be sterilized
after each use.
- Reusable semi critical items not able to withstand heat (e.g.
plastic impression trays, amalgam carriers, plastic instruments) should be processed
with high-level disinfection.
- Non-critical patient-care items that come in contact with unbroken skin (e.g. radiograph
head/cone, blood pressure cuff, facebow, pulse oximeter) in many cases need cleaning,
or if visibly soiled then cleaning followed by disinfection is adequate (low level
disinfection).
- If a non-critical item is spattered with blood or touched with a contaminated glove
or hand, it should be cleaned and disinfected.
- Non-critical items can also be protected with barriers. If contamination does occur,
just remove the barrier and throw it away. Disinfection is not necessary.
Items To Be Sterilized or Disinfected/Cleaned
|
Critical Items |
surgical instruments, periodontal knives and scaling instruments, forceps, burs
etc.
|
Sterilize
|
|
Non- Critical Items |
radiograph head/cone, blood pressure cuff, face bow, pulse oximeter etc. |
Cleaned and Disinfected |
|
Re- usable Semi-Critical Items |
plastic impression trays, amalgam carriers, plastic instruments etc. |
Disinfected |
Sterilization and Disinfection methods
Concerns about the transmission of infectious agents, such as the Hepatitis B virus
(HBV) and the Human Immuno deficiency Virus (HIV) have caused us all to be more
aware of the need to adequately sterilize and disinfect instruments and other equipments
to protect ourselves and our patients.
Sterilization Methods
A number of sterilization methods are available for heat tolerant dental instruments.
These include use of steam under pressure (steam autoclave), dry heat or unsaturated
chemical vapor. The duration of sterilization, temperatures and other operating
parameters recommended by the equipment manufacturer should be used. Additionally,
instructions for use of correct containers, wraps and chemical or biological indicators
should always be followed.
Heat sterilization methods (steam autoclave, dry heat and unsaturated chemical vapor)
are preferred for all equipment that can withstand high temperatures for several
reasons:
- Effective.
- Relatively easy to use.
- Comparatively inexpensive.
- Readily monitored for effectiveness.
Liquid chemical disinfectants/sterilants should be used only when heat will damage
an item.
Steam Sterilization
Among sterilization methods, steam sterilization is the most widely used for wrapped
and unwrapped critical and semicritical items that are not sensitive to heat and
moisture. When using an autoclave, the load must be placed so that steam can circulate
freely around each item, because steam must be able to reach all instrument surfaces
at a required temperature and pressure for a specified time in order to kill all
microorganisms and achieve sterilization. Be sure to follow the autoclave manufacturer's
operating instructions
The majority of tabletop sterilizers used in dental practice are gravity displacement
sterilizers, although prevacuum sterilizers are becoming widely available. Prevacuum
sterilizers are fitted with a pump to create a vacuum in the chamber and to ensure
air removal from the sterilizing chamber before the chamber is pressurized with
steam. Relative to gravity displacement, this procedure allows faster and more positive
steam penetration throughout the entire load.
Autoclave
An autoclave is a self locking machine that sterilizes with steam under pressure,
achieved by the high temperature. Autoclaves are the universally accepted means
for sterilization.
It is generally accepted that an autoclave chamber must reach at least 121°C at
15 Psi for a minimum of 30 minutes to ensure adequate sterilization. Sterilization
time may vary depending on the quantity and density of items in the autoclave chamber.
Overloading must be avoided.
Instruments and materials for sterilizing in the autoclave are usually enclosed
in the muslin wrappers as surgical packs. These packs should be porous to allow
steam to penetrate and reach the instruments.
The autoclave is employed for the sterilization of the instruments, extraction forceps,
surgical instruments, explorers etc. The sterilized instruments should remain wrapped
until next used.
Testing of efficacy
Autoclave indicator tape and autoclave indicator bags change colour when the proper
temperature has been reached. While the indicator tape quick-check should be conducted
with every load, it must be supplemented periodically by use of a biological indicator
(such as Bacillus spores) buried in the center of a load to confirm the sterilization.
Advantages of steam sterilization
- Is quick and easy to use.
- Allows loads to be packaged, making it easier
to maintain items in a sterile state.
- Penetrates fabric and paper wrappings.
- Can be readily monitored for effectiveness.
- Is economical and very reliable.
Disadvantages of steam heat sterilization
- May cause rust and corrosion (corrosion inhibitors such as sodium nitrite are available
that may reduce this problem).
- May damage plastics.
- May blunt certain sharp items.
Dry Heat Sterilization
Dry heat is used to sterilize materials that might be damaged by moist heat (e.g.
burs and certain orthodontic instruments). Although dry heat has the advantages
of low operating costs and being non-corrosive, it is a prolonged process and the
high temperatures required are not suitable for certain patient care items and devices
with temperatures ranging from 300 degrees F (149 degrees C) and upward can be used
for sterilization.
Dry heat sterilizers used in dentistry include static-air and forced-air types.
The static-air type is commonly called an oven-type sterilizer. The forced-air type
is also known as a rapid heat transfer sterilizer. Heated air is circulated through
the chamber at a high velocity, permitting more rapid transfer of energy from the
air to the instruments, thereby reducing the time needed for sterilization compared
to the oven-type sterilizer.
Advantages of dry heat sterilization
- Is very reliable.
- Rust and corrosion are not a problem, provided that items are dry prior to sterilization.
Is easy to use and requires little maintenance.
- Can be readily monitored for effectiveness.
Disadvantages of dry heat sterilization
- Usually requires longer processing times than do steam sterilization or unsaturated
chemical vapour.
- Damages some plastics.
- Requires careful loading.
- May char fabric.
- High temperatures may prohibit use with some materials and may melt or destroy some
metal or solder joints.
Unsaturated chemical vapour sterilization
Unsaturated chemical vapour sterilization involves heating a chemical solution of
primarily alcohol with formaldehyde in a pressurized chamber. This method of sterilization
is ideally suited to carbon steel instruments (e.g. dental burs) because the low
level of water present during the cycle results in less corrosion than might be
expected with steam sterilization. Instruments must be dry before sterilization.
Follow the manufacturer's instructions.
Advantages of chemical vapour sterilization
- Is relatively quick.
- Does not rust or corrode metal items. Is very reliable.
- Can be used with packaged items (paper packaging only).
- Can be monitored for effectiveness.
Disadvantages of chemical vapour sterilization
- Requires good ventilation owing to fumes. Won't penetrate fabric- wrapped packs.
Damages some plastics.
- Requires replacement of special solution, increasing cost.
- Requires hazardous waste disposal of the sterilizing solution.
Sterilization of Unwrapped Instruments
An unwrapped sterilization cycle (sometimes called flash-sterilization) is a method
of sterilizing patient care items for immediate use. Critical and semi- critical
items that have been sterilized unwrapped should be transferred immediately, using
aseptic technique for the ultimate use. The unwrapped sterilization cycle in tabletop
sterilizers is usually preprogrammed by the manufacturer to a specific time and
temperature setting.
Thorough cleaning and drying of instruments precedes the unwrapped cycle. Mechanical
monitors are checked and chemical indicators are used for each cycle. Care is taken
to avoid thermal injury to dental workers and patients.
Items are handled and transported aseptically to the point of use to maintain sterility.
As implantable devices should be quarantined after sterilization until the results
of biological monitoring are known, unwrapped or flash sterilization of implantable
devices is not recommended.
Other Methods of Sterilization
Liquid chemical germicides: Heat-sensitive critical and semicritical instruments
and devices can be sterilized by immersing them in liquid chemical germicides. However,
items sterilized in this manner can require approximately 12 hours of complete immersion.
Additionally items sterilized in this manner must be rinsed with sterile water to
remove any toxic or irritating residues, handled using sterile gloves and dried
with sterile towels, delivered to the point of use in an aseptic manner and then
used immediately.
Because of these limitations, they are almost never used to sterilize instruments.
Rather, these chemicals are more often used for high-level disinfection of heat-sensitive
semicritical instruments and devices. Shorter immersion times (12-90 minutes) make
high-level disinfection more practical than sterilization; however, instruments
and devices disinfected in this manner must still be handled as if sterile (e.g.
rinsed with sterile water, dried with sterile towels, etc.) and used immediately.
Chemical sterilants (e.g. glutaraldehyde, peracetic acid, hydrogen peroxide) are
powerful, sporicidal chemicals and are highly toxic. Instruments can be sterilized
by placing them in a 2% solution of glutaraldehyde for 6-10 hours. The chemical
sterilants must only be used according to manufacturer's instructions and for those
applications indicated on their label. Misapplications include use as an environmental
surface disinfectant or instrument holding solution.
Alcohols are effective as skin antiseptics.Usually a 50% to 80% ethyl alcohol solution
is recommended.
In general, use of heat-sensitive semi- critical items that must be processed with
liquid chemical germicides is discouraged; heat-tolerant or disposable alternatives
are available for the majority of such items.
Ethylene oxide gas (ETG): Low temperature sterilization with ETG has been extensively
used in larger health care facilities. However, extended sterilization times of
10-48 hours and potential hazards to patients and the dental team that require stringent
health and safety requirements make this method impractical for the private-practice
setting.
Bead sterilizers: Historically, bead sterilizers have been used in dentistry to
sterilize small metallic instruments (e.g. endodontic files). This method employs
a heat transfer device. The media used are glass beads or salt and the temperature
achieved is 220°C.
The method empolys submersion of small instruments such as endodontic files and
burs into the beads; they are sterilized in 10 seconds provided they are clean.
Sterilization and Disinfection of Patient Care Items
The intended use of the patient-care item should determine the recommended level
of disinfection. Dental practices should follow the product manufacturer's directions
regarding concentrations and exposure time for disinfectant activity relative to
the surface to be disinfected.
Transporting and Processing Contaminated Critical and Semicritical
Patient-Care Items
- Dental practitioners can be exposed to micro-organisms on contaminated instruments
and devices through percutaneous injury, contact with non-intact skin on the hands
or contact with mucous membranes of the eyes, nose or mouth.
- Contaminated instruments should be handled carefully to prevent exposure to sharp
instruments that can cause a percutaneous injury.
- Sterilization is a complex process requiring specialized equipment, adequate space,
qualified dental practitioner who are provided with ongoing training and regular
monitoring for quality assurance.
-
Correct cleaning, packaging, sterilizer loading procedures, sterilization methods
or high-level disinfection methods should be followed to ensure that an instrument
is adequately processed and safe for reuse on patients.
Instrument Processing Area
- Dental practitioners should process all instruments in a designated central processing
area to ensure quality control and safety. The central processing area should be
divided into sections for
- Receiving, cleaning and decontamination.
- Preparation and packaging.
-
Sterilization.
- Storage.
- Ideally, walls or partitions should separate the sections to control traffic flow
and contain contaminants during processing.
- When physical separation of these sections cannot be achieved, adequate spatial
separation is necessary to prevent contamination of clean areas.
- Space should
be adequate for the anticipated volume of work and storage of items.
Receiving, Cleaning and Decontamination
- Reusable instruments, supplies and equipment should be received, sorted, cleaned
and decontaminated in one section of the processing area.
- Cleaning should precede
all disinfection and sterilization processes; it will remove organic and inorganic
contamination.
- Removal of debris and contamination is achieved either by scrubbing with a surfactant,
detergent and water or by an automated process (e.g. ultrasonic cleaner or washer-
disinfector) using chemical agents.
- If visible debris, whether inorganic or
organic matter, is not removed, it will interfere with microbial inactivation and
can compromise the disinfection or sterilization process.
- After cleaning, instruments should be rinsed with water to remove chemical or detergent
residue.
- Splashing should be minimised during cleaning and rinsing.
- Before final disinfection or sterilization, instruments should be handled as though
contaminated.
- Considerations in selecting cleaning methods and equipment include
:-
- Efficacy of the method, process and equipment.
- Compatibility with items
to be cleaned.
- Occupational health and exposure risks.
- Use of automated cleaning equipment (e.g. ultrasonic cleaner or washer- disinfector)
does not require presoaking or scrubbing of instruments and can increase productivity,
improve cleaning effectiveness and decrease worker exposure to blood and body fluids.
- Using automated equipment can be safer and more efficient than manually cleaning
contaminated instruments.
- If manual cleaning is not performed immediately, place
instruments in a puncture-resistant container and soak them with disinfectant/detergent
or an enzymatic cleaner to prevent drying of contaminated debri and makes cleaning
easier and less time-consuming.
- It is recommended not to use liquid chemical sterilant/high-level disinfectant (e.g.
glutaraldehyde) as a holding solution.
- Do not scrub sharp instruments.
- Wear puncture-resistant, heavy-duty utility gloves when handling or manually cleaning
contaminated instruments and devices.
- Do not blindly reach into trays or containers
holding sharp instruments. (e.g. sinks filled with soapy water in which sharp instruments
have been soaked).
- Use strainer-type basket to hold instruments and forceps. Wear a mask, protective
eyewear or face shield, gown or jacket to avoid splashing.
Preparation and Packaging
- In another section of the processing area, cleaned instruments and other dental
supplies should be inspected, assembled into sets or trays and wrapped, packaged,
or placed into container systems for sterilization.
- Hinged instruments should
be processed open and unlocked.
- An internal chemical indicator should be placed in every package.
- In addition, an external chemical indicator (e.g. chemical indicator tape) should
be used when the internal indicator cannot be seen from outside the package.
-
Dental practices should refer to the manufacturer's instructions regarding use and
correct placement of chemical indicators.
- Critical and semicritical instruments should be wrapped or placed in containers
(e.g. cassettes or organizing trays) designed to maintain sterility.
- Packaging
materials (e.g. wraps or container systems) allow penetration of the sterilization
agent and maintain sterility of the processed item after sterilization.
- Maintain
sterility of instruments during transport and storage by using wrapped perforated
instrument cassettes, peel pouches of plastic or paper and sterilization wraps (i.e.
woven and non woven).
- Packaging materials should be designed for the type of sterilization process being
used.
Sterilization of Unwrapped Instruments
- Clean and dry instruments prior to the unwrapped sterilization cycle.
- Use
mechanical and chemical (place an internal chemical indicator among the instruments
or items to be sterilized) indicators for each unwrapped sterilization cycle.
-
Allow unwrapped instruments to dry and cool in the sterilizer before they are handled
to avoid contamination and thermal injury.
- Semi- critical instruments that will be used immediately or within a short time
can be sterilized unwrapped on a tray or in a container system, provided that the
instruments are handled aseptically during removal from the sterilizer and transported
to the point of use.
- Critical instruments intended for immediate reuse can be sterilized unwrapped provided
that the instruments are maintained sterile during removal from the sterilizer and
transported to the point of use.
- Do not sterilize implantable devices unwrapped.
- Do not store critical instruments unwrapped.
Sterilization Monitoring
Use mechanical, chemical and biological monitors according to the manufacturer's
instructions to ensure the effectiveness of the sterilization process.
Monitor each load with mechanical (e.g. time, temperature and pressure) and chemical
indicators. Place a chemical indicator on the inside of each package. If the internal
indicator is not visible from the outside, also place an exterior chemical indicator
on the package. Do not use instrument packs if mechanical or chemical indicators
indicate inadequate processing.
Monitor sterilizers at least weekly by using a biological indicator with a matching
control (i.e. biological indicator and control from same lot number). Use a biological
indicator for every sterilizer load that contains an implantable device. Verify
results before using the implantable device, whenever possible.
The following are recommended in the case of a positive spore
test:
- Remove the sterilizer from service and review sterilization procedures (e.g. work
practices and use of mechanical and chemical indicators) to determine whether operator
error could be responsible;
- Retest the sterilizer by using biological, mechanical and chemical indicators after
correcting any identified procedural problems and
- If the repeat spore test is negative and mechanical and chemical indicators are
within normal limits, put the sterilizer back in service.
The following are recommended if the repeat spore test is
positive:
- Do not use the sterilizer until it has been inspected or repaired or the exact reason
for the positive test has been determined,
- Recall, to the extent possible and reprocess all items processed since the last
negative spore test and
- Frequently clean all working areas contaminated with latex powder or dust.
-
Have emergency treatment kits with latex-free products available at all times.
-
Before placing the sterilizer back in service, rechallenge the sterilizer with biological
indicator tests in three consecutive empty chamber sterilization cycles after the
cause of the sterilizer failure has been determined and corrected.
Maintain sterilization records (i.e. mechanical, chemical and biological).
Sterilization Process
Storage Of Sterilized Items and Clean Dental Supplies
- The storage area should contain enclosed storage for sterile items and disposable
(single-use) items.
- Storage practices for wrapped sterilized instruments can be either date or event-
related.
- Packages containing sterile supplies should be inspected before use to verify barrier
integrity and dryness.
- Ideally a product should remain sterile indefinitely, unless an event causes it
to become contaminated (e.g. torn or wet packaging).
- Even for event-related
packaging, minimally, the date of sterilization should be placed on the package
and if multiple sterilizers are used in the facility, the sterilizer used should
be indicated on the outside of the packaging material to facilitate the retrieval
of processed items in the event of a sterilization failure.
- If packaging is compromised, the instruments should be recleaned, packaged in new
wrap and sterilized again.
- Clean supplies and instruments should be stored in
closed or covered cabinets, if possible.
- Dental supplies and instruments should not be stored under sinks or in other locations
where they might become wet.
Surface Infection Control
In the dental office, the surface contamination can occur during patient care. Certain
surfaces especially the ones touched regularly during patient care (e.g. light handles,
unit switches and drawer knobs can serve as reservoirs of microbial contamination).
These surfaces can be divided into clinical contact surfaces and house keeping surfaces.
Clinical Contact Surface
Clinical contact surfaces can be directly contaminated from patient materials either
by direct spray or spatter generated during dental procedures or by contact with
dental professionals gloved hands. These surfaces can subsequently contaminate other
instruments, devices, hands or gloves.
- Examples of such surfaces include:-
- light handles,
- switches,
- dental radiograph equipment,
- dental chairside computers,
- reusable containers of dental materials,
- drawer handles,
- faucet handles,
- countertops,
- pens,
- telephones and
- doorknobs.
- Barrier protection of surfaces and equipment can prevent contamination of surfaces
which are difficult to clean. Barriers includes using clear plastic wrap, bags,
sheets, tubing and plastic-backed paper or other materials impervious to moisture.
- These coverings can become contaminated. They should be removed and discarded between
patients, while DHCP are still gloved. After removing the barrier, examine the surface
to make sure it did not become soiled. The surface needs to be cleaned and disinfected
only if contamination is evident. Otherwise, after removing gloves and performing
hand hygiene, DHCP should place clean barriers on these surfaces before the next
patient.
- If barriers are not used, surfaces should be cleaned and disinfected between patients
by using a disinfectant with an HIV, HBV claim (i.e. low- level disinfectant) or
a tuberculocidal claim (i.e. intermediate-level disinfectant). Intermediate-level
disinfectant should be used when the surface is visibly contaminated with blood
or OPIM.
- General cleaning and disinfection are recommended for surfaces, dental unit surfaces
and countertops at the end of daily activities and these may have become contaminated
since their last cleaning. To facilitate daily cleaning, treatment areas should
be kept free of unnecessary equipment and supplies.
- Manufacturers of dental devices and equipment should provide information regarding
material compatibility with liquid chemical germicides, whether equipment can be
safely immersed for cleaning and how it should be decontaminated if servicing is
required.
- Because of the risks associated with exposure to chemical disinfectants and contaminated
surfaces, dental professionals who perform environmental cleaning and disinfection
should wear gloves and other PPE to prevent occupational exposure to infectious
agents and hazardous chemicals.
- Chemical and puncture-resistant utility gloves offer more protection than patient
examination gloves when using hazardous chemicals.
Housekeeping Surfaces
- Evidence does not support that housekeeping surfaces (e.g. floors, walls and sinks)
pose a risk for disease transmission in dental health- care settings.
- These need to be cleaned only with a detergent and water or a disinfectant/detergent,
depending on the nature of the surface and the type and degree of contamination.
- Schedules and methods vary according to the area (e.g. dental operatory, laboratory,
bathrooms, or reception rooms), surface, amount and type of contamination.
- Floors should be cleaned regularly and spills should be cleaned up promptly. A disinfectant/detergent
designed for general housekeeping purposes should be used in patient-care areas
if uncertainty exists regarding the nature of the soil on the surface (e.g. blood
or body fluid contamination versus routine dust or dirt).
- Unless contamination is reasonably anticipated or apparent, cleaning or disinfecting
walls, window drapes and other vertical surfaces is unnecessary.
- However, when housekeeping surfaces are visibly contaminated by blood or OPIM, prompt
removal and surface disinfection is essential.
- Part of the cleaning strategy is to minimize contamination of cleaning solutions
and cleaning tools (e.g. mop heads or cleaning cloths).
- Mops and cloths should be cleaned after use and allowed to dry before reuse or single-use,
disposable mop heads and cloths should be used to avoid spreading contamination.
Disinfection of Dental Prosthesis and Impressions
To protect the laboratory personnel, impressions, gypsum casts, fixed and removable
prosthesis, wax rims and bite registration records should be appropriately disinfected
before sending them to the dental laboratory.
Dental prosthesis, impressions, orthodontic appliances and other prosthodontic materials
(e.g. occlusal rims, temporary prosthesis, bite registrations) should be thoroughly
rinsed under gentle running water to remove blood, saliva or debris.
The disinfection can be done by a short-term immersion in 0.5% or 1% sodium hypochlorite.
In addition, immersion in glutaraldehydes, povidone-iodine diluted in water or halogenated
phenol has no apparent effect on the dimensional stability of rubber materials.
The best time to clean and disinfect the impressions, prosthesis or appliances is
as soon as possible after removal from the patients mouth, before drying of blood
and saliva.
Extracted Teeth and Biopsy Specimens
Handling of Biopsy specimens
- To protect persons handling and transporting biopsy specimens, each specimen must
be placed in a sturdy, leakproof container with a secure lid for transportation.
- Care should be taken when collecting the specimen to avoid contaminating the outside
of the container.
- If the outside of the container becomes visibly contaminated, it should be cleaned
and disinfected or placed in an impervious bag.
- The container must be labeled
with the biohazard symbol during storage, transport, shipment and disposal.
Handling of Extracted Teeth The extracted teeth
should either be disposed or should be used for educational purpose.
-
Disposal
- Extracted teeth should be considered as potentially infectious material and should
be disposed in medical waste containers.
- Extracted teeth sent to a dental laboratory for shade or size comparisons should
be cleaned, surface- disinfected with a disinfectant with intermediate- level activity.
- Extracted teeth containing dental amalgam should not be placed in a medical waste
container meant for incineration as the final disposal.
-
Educational Settings
- Extracted teeth are occasionally collected for use in educational training.
- These teeth should be cleaned of visible blood and gross debris and maintained in
a hydrated state in a well- constructed closed container during transport.
- Because these teeth will be autoclaved before clinical exercises or study, use of
the most economical storage solution (e.g. water or saline) might be practical.
- Liquid chemical germicides can also be used to disinfect both external surface and
interior pulp tissue.
- Before being used in an educational setting, the teeth should be heat- sterilized
to allow safe handling. Microbial growth can be eliminated by using an autoclave
cycle for 40 minutes.
- Use of teeth that do not contain amalgam is preferred because they can be safely
autoclaved.
- Extracted teeth containing amalgam restorations should not be heat- sterilized because
of the potential health hazard from mercury vaporization and exposure.
- If extracted
teeth containing amalgam restorations are to be used, immersion in 10% formalin
solution for 2 weeks should be effective in disinfecting both the internal and external
structures of the teeth.
Cleaning and Disinfection Strategies for Blood Spills
The majority of blood contamination in dentistry result from spatter during dental
procedures using rotary or ultrasonic instrumentation. Although no evidence supports
that HBV, HCV or HIV have been transmitted from a housekeeping surface, prompt removal
and surface disinfection of an area contaminated by either blood or OPIM are appropriate
infection-control practices.
- The person assigned to clean the spill should wear gloves and other PPE as needed.
-
Visible organic material should be removed with an absorbent material (e.g. disposable
paper towels discarded in a leak-proof, appropriately labeled container).
- Non-porous surfaces should be cleaned and then decontaminated with effective disinfectant.
-
1:100 dilution of sodium hypochlorite (e.g. approximately ¼ cup of 5.25% household
chlorine bleach to 1 gallon of water) is an inexpensive and effective disinfecting
agent.
Carpeting and Cloth Furnishings
Carpeting is more difficult to clean than non- porous hard- surface flooring and
it cannot be reliably disinfected, especially after spills of blood and body substances.
Cloth furnishings pose similar contamination risks in areas of direct patient care
and places where contaminated materials are managed (e.g. dental operatory, laboratory
or instrument processing areas). For these reasons, use of carpeted flooring and
fabric-upholstered furnishings in these areas should be avoided.
Infection Control for Water Unit
Dental unit waterlines (i.e. narrow-bore plastic tubing that carries water to the
high-speed handpiece, air/water syringe and ultrasonic scaler) may become colonized
with micro-organisms, including bacteria, fungi and protozoa. These micro-organisms
colonize and replicate on the interior surfaces of the waterline tubing and form
a biofilm, which serves as a reservoir that can amplify the number of free-floating
(i.e. planktonic) micro-rganisms in water used for dental treatment and exhibit
limited pathogenic potential for immunocompetent persons.
Patient material (e.g. oral microorganisms, blood and saliva) can enter the dental
water system during patient treatment hence it is necessary to flush out the patient
material that might have entered the turbine, air and waterlines.
Strategies To Improve Dental Unit Water Quality
- Dental waterlines should be flushed at the beginning of the clinic day to reduce
the microbial load.
- However, this does not affect biofilm in the waterlines or reliably improve the
quality of water used during dental treatment as the recommended value of <500 CFU/mL
cannot be achieved by using this method.
- All incoming waterlines from the public water system inside the dental office (e.g.
faucets, waterlines and dental equipment) should be flushed for 1 to 5 minutes .
- The length of time needed can vary with the type and length of the plumbing system
of the office. After the incoming public water system lines are flushed, dental
unit waterlines should be disinfected according to the manufacturer's instructions.
-
Quality of water used can be improved by including self- contained water systems
combined with chemical treatment, in-line microfilters and combinations of these
treatments.
- Simply using source water containing <500 CFU/mL of bacteria (e.g. tap, distilled
or sterile water) in a self- contained water system will not eliminate bacterial
contamination in treatment water if biofilms in the water system are not controlled.
- Removal or inactivation of dental waterline biofilms requires use of chemical germicides.
- Physically flush out patient material that might have entered the turbine, air or
waterlines by operating it to discharge water and air for a minimum of 20-30 seconds
after each patient.
- Users should consult the owner's manual or contact the manufacturer to determine
whether testing or maintenance of anti retraction valves or other devices is required.
Maintenance and Monitoring of Dental Unit Water
- Dental practitioners should be trained regarding water quality, biofilm formation,
water treatment methods and appropriate maintenance protocols for water delivery
systems.
- Water treatment and monitoring products require strict adherence to maintenance
protocols and noncompliance of treatment regimens is associated with persistence
of microbial contamination in treatment systems.
Handpieces and Other Devices
Multiple semicritical dental devices that touch mucous membranes are attached to
the air or waterlines of the dental unit. Among these devices are high and low-speed
handpieces, prophylaxis angles, ultrasonic and sonic scaling tips, air abrasion
devices and air and water syringe tips. These devices have the potential for retracting
oral fluids into their internal compartments. Restricted physical access limits
their cleaning. This indicates that retained patient material can be expelled intraorally
during subsequent uses.
- Any dental device connected to the dental air/water system that enters the patient's
mouth should be run to discharge water, air or a combination for a minimum of 20-
30 seconds after each patient.
- This procedure physically flushes out patient material that might have entered the
turbine, air and waterlines.
- Heat methods can sterilize dental handpieces and
other intraoral devices attached to air or waterlines.
- Proper scrubbing with detergent, water and drying followed by wiping with a suitable
chemical disinfectant is essential for those ultrasonic scalers, handpieces and
air syringes that cannot be sterilized.
- For processing any dental device that
can be removed from the dental unit air or waterlines, neither surface disinfection
nor immersion in chemical germicides is an acceptable method.
- Manufacturer's instructions for cleaning, lubrication and sterilization should be
followed to ensure both the effectiveness of the process and the longevity of handpieces.
- Some components of dental instruments are permanently attached to dental unit waterlines
and although they do not enter the patient's oral cavity, they are likely to become
contaminated with oral fluids during treatment procedures.
- Components (e.g. handles or dental unit attachments of saliva ejectors, high- speed
air evacuators and air/water syringes) should be covered with impervious barriers
that can be changed after each use.
- If the item becomes visibly contaminated during use, dental practitioner should
clean and disinfect with a disinfectant (intermediate- level) before use on the
next patient.
Saliva Ejectors
- Backflow from low-volume saliva ejectors occurs when the pressure in the patient's
mouth is less than that in the evacuator.
- Backflow in low-volume suction lines can occur and micro-organisms present in the
lines retracted into the patient's mouth when a seal around the saliva ejector is
created (e.g. by a patient closing lips around the tip of the ejector, creating
a partial vacuum).
- This backflow can be a potential source of cross- contamination.
- Although no adverse health effects associated with the saliva ejector have been
reported, practitioners should be aware that in certain situations, backflow can
occur when using a saliva ejector.
Single-Use or Disposable Devices
- These must not be carried in uniform or clothing pockets.
- It is designed to be used on one patient and then discarded, not reprocessed for
use on another patient (e.g. cleaned, disinfected, or sterilized).
- It is not heat-tolerant and cannot be reliably cleaned. Examples include syringe
needles, prophylaxis cups and brushes and plastic orthodontic brackets.
- Certain
items (e.g. prophylaxis angles, saliva ejectors, high- volume evacuator tips and
air/water syringe tips) are commonly available in a disposable form.
- These items (e.g. cotton rolls, gauze and irrigating syringes) should be sterile
at the time of oral surgical procedures.
- Because of the physical construction of certain devices (e.g. burs, endodontic files
and broaches) cleaning can be difficult.
- In addition, deterioration can occur
on the cutting surfaces of some carbide/diamond burs and endodontic files during
processing and after repeated processing cycles, leading to potential breakage during
patient treatment.
- These factors, coupled with the knowledge that burs and endodontic instruments exhibit
signs of wear during normal use, should be considered as single-use devices.
Checklist
- Flush devices connected to dental unit air and water system for 20- 30 seconds between
patients.
- Flush dental waterlines at the beginning and at the end of the day.
- Insert only sterile needles and syringes into multi-dose vials.
- Clean and disinfect blood spills immediately.
- Dispose of all single-use
devices after use on patient.
- Use sterilized instruments for every patient.
- Decontaminate or properly contain and label extracted teeth.
- Properly dispose of extracted teeth containing amalgam.
- Heat sterilize extracted
teeth before using them in educational settings.
- Disinfect dental prostesis
and impressions before sending to the laboratory.