Introduction
Prof. Roentgen discovered X–rays in 1895. These are used in medicine, research and
industry. Medical X-rays were first used in India in 1898 within three years of
discovery of X-rays. Consequently X-rays were also used in dentistry. Radiographs
provide an insight on the fourth dimension that cannot be seen clinically.
Despite the inherent hazards, theri utility was more in comparisons to risks. This,
coupled with the increasing use of radiology in industries, led to the formulation
of the ‘Atomic Energy Act, 1962,’ to ensure radiation safety in the use of radiation
generating plants Under Section 17 of the Act. Pursuant to the provisions of the
Act, the Central Government had promulgated the Radiation Protection Rules, 1971,
which stipulate basic safety standards for all types of radiation application in
medicine, industry, research etc. Appropriate radiation surveillance procedures
have been issued under Rule 15 to ensure radiation protection in various types of
applications.
Practitioners who administer ionizing radiation must be familiar with the magnitude
of radiation exposure encountered in medicine and dentistry, the possible risk that
such exposure entails and the methods that can be used to attenuate exposure and
decrease dosage. This information provides the necessary background to explain to
the concerned patients the benefits and possible hazards involved in the use of
X-Rays.
Indications of Radiology
Numerous conditions of the teeth and jaws may not show clinical signs or symptoms
but can be, detected with the help of dental radiographs. Some of the common diseases,
lesions, and conditions revealed on dental radiographs are:
- Missing teeth
- Supernumerary teeth
- Dental caries
- Impacted
teeth
- Periodontal disease
- Tooth abnormalities
- Retained roots
-
Cysts and tumors
Guidelines for Ordering Radiographs
- Diagnostic
- Administrative
- Medico-legal
- Industry
- Research
Diagnostic
Intra-Oral Radiographs
Peri-apical
Bitewing
Occlusal
Extra-Oral Radiographs
Lateral Oblique
Panoramic
P A Mandible
Sub-Mento Vertex
Sub-Mento Vertex
Waters view
T M J Projections; etc.
Administrative
Means that the exposed individual receives no benefits from the exposure.
Medico-legal X-rays may be taken for reasons other than diagnosis like Insurance,
civil litigation, weapons or drug search, immigration and emigration, prisoners,
pre employment etc
X-rays are used for non destructive testing and have a wide range of application
in industries. They are used to detect crack in metal structures - the body of aeroplanes
and automobiles, the quality of welding in moulds and metal castings, the presence
of pearls in oysters etc
X-rays are used to study the structure of crystals, which is called X-rays crystallography.
Also used in pharmaceuticals, lasers etc.
Responsibilities of the dental faculty
The referring dental surgeon should provide a clear request describing the patient’s
problem and indicating the clinical objectives, so that the radiologist can carry
out the correct X-ray examination.
Before prescribing an X-ray examination the referring dental surgeon should be satisfied
that the necessary information is not available, either from radiographic examinations
already done, or from any other medical tests or investigations.
If two or more dental imaging procedures are available and give the desired diagnostic
information, then the procedure that presents the least overall risk to the patient
should be chosen.
Dose Equivalent Limits
Receiving 50 mSv of whole body radiation exposure in 1 year as a result of performing
one’s occupation is considered to present minimal risk. However every effort should
be made to keep the dose to all individuals as low as possible. All unnecessary
radiation exposure should be avoided. This is a philosophy of radiation protection
everyone should recognize. It is based on the ALARA principle (As Low As Reasonably
Achievable), which recognizes the possibility that no matter how small the dose,
some adverse health effect may occur.
Radiation Dose From Diagnostic Radiology
Principal sources of radiation dose for members of the public are natural background
radiation and the medical / dental applications of radiation.
Absorbed Dose In The Body Tissues
Absorbed dose in body tissues are used to indicate the risk to specific tissues.
The total risk from any particular X-ray examination should ideally comprise the
risk from all the radiation sensitive tissues that are irradiated.
Heath Effects of Ionizing Radiations
The biological effect of ionizing radiation can be extremely damaging. The effects
are classified as: -
- Somatic Non – Stochastic (certainty) effects.
- Somatic Stochastic effects
- Genetic Stochastic effects
Somatic Non–Stochastic effects are predominant at high doses of radiation, while
Somatic Stochastic and Genetic Stochastic effects predominate even with low doses.
The ALARA principle should be followed.
Risk Of Neoplasia And Hereditary Effects
|
Effect
|
Nominal risk per milligray
|
Nominal Ratios Total Cancers Fatal Cancers
|
|
Hereditary (General)
|
1 in 2,50,000
|
|
|
Cancers
|
Fatal
|
|
|
Leukemia (Active bone marrow)
|
1 in 5,00,000
|
1.05
|
|
Breast cancer (Females)
|
1 in 2,00,000
|
1.6
|
|
Lung cancer
|
1 in 5,00,000
|
1.05
|
|
Thyroid cancer
|
1 in 20,00,000
|
21
|
|
Others (Combined)
|
1 in 20,00,000
|
1.3
|
|
|
|
|
|
Irradiation In Utero
|
|
|
|
|
|
|
|
TIME AFTER CONCEPTION
|
NOMINAL RISK PER MILIGRAY
|
|
|
|
|
|
|
First two weeks
|
Minimal
|
|
|
3rd through 8th week
|
Potential for malformation of organs
|
|
|
8th through 15th week
|
Severe mental retardation (1 in 2,500)
|
|
|
15th through 25th week
|
Severe mental retardation (1 in 10,000)
|
|
|
Throughout pregnancy
|
Childhood cancer (1 in 50,000)
|
|
Examination of Women Having Reproductive Capacity & During Pregnancy
Because of radiation risk to an embryo, the possibility of a woman being pregnant
has to be considered in deciding whether to conduct an examination that might cause
irradiation to the lower abdomen.
During the first 10 days following the onset of menstrual period, there is minimum
radiation risk since no conception will have occurred. The radiation risks to a
child who had been irradiated in-utero during the remainder of the first month following
the onset of menstruation (i.e. during approx. the first two weeks after conception
is likely to be so small that there need be no special limitation on X-ray examinations).
Period of 18-55 days after conception is critical to organ formation. Use of lead
aprons is mandatory on such patients.
Patient Protection
X-rays cause biologic changes in living cells and adversely affects all living tissues.
With the use of proper patient protection techniques, the amount of X-rays received
by the patient can be minimised. Patient protection techniques can be used prior
to, during, and after X-ray exposure.
Patient protection measures can be employed prior to any X-ray exposure by:
1) Appropriate prescribing of dental radiographs
2) Proper equipment
The dental X-ray tube head must be equipped with appropriate aluminum filters, lead
collimator and position-indicating device.
3) Size Of X-Ray Field
- Smallest practicable X-ray field
- Accurate positioning
i. Total radiation
energy delivered to patient is reduced
- Reduction of scattered radiation
i. This improves imaging quality.
- Use of beam limiting devices (collimation)
4) Collimation
A collimator may have either a round or rectangular opening. When using a circular
collimator, diameter of the X-ray beam of no more than 2.75 inches should be present
as it exits from the position-indicating device and reaches the skin of the patient.
5) Distance From The Focal Spot To The Skin Or Image Receptor
There are two techniques for taking radiographs depending on the focal spot to film
distance
- Short cone technique (8” distance)
- Long cone technique (16” distance)
Longer focal spot to image receptor distances have clinical advantages. Therefore
long cone is preferable.
6) Position-Indicating Device: There are three basic types of PIDs
The open-ended, lead-lined rectangular or round PIDs that do not produce scatter
radiation are used. The long PID is preferred because less divergence of the X-ray
beam occurs.
7) Total Filtration In The X-Ray Beam
For dental equipment with X-ray tube voltages not exceeding 70 kVp the total permanent
filtration in the X-ray beam should be equivalent to not less than 1.5 mm of aluminum.
At higher X-ray tube voltages the total filtration should be equivalent to not less
than 2.5 mm of aluminum of which 1.5 mm should be permanent.
Patient protection measures are used not only prior to X-ray exposure but also during
exposure.
1) Control Of Irradiation & Recording Of Time
- Operating switches should be constructed such that irradiation can be terminated
manually at any time.
- Also, the spring loaded type “Deadman” switches should
be cautiously used to avoid accidental exposure
2) Thyroid Collar
Radiation dose to the thyroid is considered the largest component of the effective
dose in dental radiography.
3) Lead Apron
Although scatter radiation to the patient's abdomen is extremely low, lead aprons
should be used to minimize patient's exposure to radiation.
4) Fast Film
The use of faster films (E- or F-speed) is preferred because they reduce the radiation
dose by more than 50 percent compared with D-speed film.
5) Film-Holding Devices
Receptor / Film holders that position the receptor to coincide with the collimated
X-ray beam should be used.
6) Exposure Factor Selection
The dental radiologist cannot control all the exposure factors as the kilovoltage
peak (kVp), milliamperage are preset by the manufacturer, but the time settings
on the control panel can be adjusted as per the requirement. A setting of 65 to
70 kVp and 7 mA to 10 mA is the range available in dental intra oral machines.
7) Intensifying Screens & Films
Screens containing high efficiency materials require less radiation than conventional
ones to provide similar image quality. The materials used are rare earths, samarium,
barium, tantalum, terbium, etc.
8) Proper Technique
Proper technique helps to ensure the diagnostic quality of films and reduce the
amount of exposure a patient receives and avoids the need to retake radiographs.
9) Control Of Scattered Irradiation To The Image Receptor
- The use of grids should be moderated. Stationary and moving grids are used in practice.
-
The grid ratio should be 10:1.
1) Proper Film Handling
From the time the films are exposed until they are processed, careful handling is
of the utmost importance. Artifacts due to improper film handling result in non-diagnostic
films.
2) Proper Film Processing
Correct processing techniques are necessary to give reproducible radiographs of
optimum diagnostic value with minimum dose to the patient. Film processing should
be performed under the manufacturer recommended conditions with proper processing
equipment and a darkroom with safe-lights. Alternatively, an automatic processor
with an appropriate safe light hood may be used.
3) Darkroom practices
Proper radiological darkroom practices should be followed. These include maintaining
a darkroom with adequate ventilation, avoiding repeated skin contact with processing
chemicals and avoiding microbial contamination in handling film packets. Darkrooms
should be checked routinely for light leaks.
4) Proper Reading And Interpretation Of The Resultant Radiograph
Radiographic images should be viewed with an illuminated viewer to obtain maximum
available information.
5) Quality Assurance
All radiographic images captured by using radiographic facilities should establish
quality assurance programmes whose structure and scope are determined by the needs
and complexity of each facility.
6) Personal Facility Monitoring Devices
TLD badges, electronic dosimeter.
Regulatory Controls & Operational Safety
Regulatory Controls
|
Design certification
|
The X-ray diagnostic equipment must be certified to meet the design specifications
stipulated by the ‘Atomic Energy Act.’
|
|
Registration of equipment
|
the sales, transfer, lease or loan of X-ray equipment must be registered with the
competent authority.
|
|
Report of unsafe equipment
|
The personnel servicing the X-ray installation must report the safety of the installation
to the competent authority in case of unsafe functioning.
|
|
Inspection
|
Periodic inspection of the X-ray installation must be done and duly recorded
by the dental surgeon or radiologist.
|
|
Guidelines
|
The competent authority must formulate guidelines for the standardized use of radiology
and minimize the health risk involved in the procedure.
|
|
Operational Safety Commissioning
|
When diagnostic X-ray unit is installed and after structural modifications
a Radiation Protection Survey is conducted. New units manufactured in the country
are being type approved by AERB (Atomic Energy Regulatory Board) with technical
support from BARC (Bhabha Atomic Research Centre).
|
Operation Of
Equipment
|
Primary beam is directed towards areas of minimum occupancy.
Installation of more than 1 X-ray unit in same room should be discouraged.
Only patient (whose examination is to be carried out) should be allowed in
room.
|
|
Control Panel
|
If in the X-ray unit, it must be located as far away from X-ray unit stand
as possible and duly shielded by a protective barrier.
|
|
Furnishing & Fixtures
|
Room should have only essential furniture and facilities needed for the examination
so as to discourage presence of uninvolved staff and patients in the room.
|
|
Monitoring Of Exposure To Personnel
|
TLD badges, dosimeters etc. should be available to monitor the radiation dose received
by operating personnel.
|
|
|
|
|
Top
|
|
X-Ray Room Layout
|
|
|
Location Of X-Ray Installation
|
The rooms housing diagnostic X-ray units and equipment should be located as far
away as possible from areas of high occupancy and general traffic, such as maternity
and pediatric wards.
|
|
Layout
|
Layout should aim at providing integrated facilities so that handling of equipment
and related operations can be performed with adequate protection.
The number of doors for entry to room should be minimum.
Doors and passages leading to installation should permit safe and easy transport
of equipment and patients.
Dark room should be located such that primary X-ray beam is not directed on
it.
|
|
Position And Distance Rule
|
Operator should be at least 6 ft. away from source at an angle of and 90 - 135 degrees.
|
|
Room Size
|
Must be spacious
There should be a minimum area of 150 sq. feet for dental radiography.
|
|
Shielding
|
Appropriate structural shielding should be provided for wall, ceiling and
floor so that maximum permissible dose i.e. 50 mSv for occupational workers and
1 mSv for public should not be exceeded.
9-inch brick wall.
Lead barrier
|
|
Opening & Ventilation
|
Unshielded openings if provided in an X-ray room for ventilation or natural light
etc. must be located at a height of 6 ft. above finish floor level outside the x-ray
room.
|
|
Illumination Control
|
Safelights (red light) 1.2 m away from work surface with 15 watts bulb and
filters suitable for the types of films being used.
|
|
Equipment Layout
|
Installed in a way that in normal use the beam is not directed towards
control panel, doors, windows or areas of high occupancy.
1 mm of lead sheet protection for entrance door.
1.5 mm lead barrier should be used.
Lead glass for the windows should be used
|
|
Control Panel
|
Control panel should be contiguous
Stretchable cable length 2 mts. (at least) should be provided
For higher than 125 kV control panel should be in separate room
Appropriate shielding
|
|
Waiting Areas
|
Should be outside the X-ray room.
|
Warning Light &
Placard
|
Warning signal (red light) and a radiation warning sign at a conspicuous place outside
room must be kept on when unit is in use and placard should be displayed prominently.
|
Schedule for Radiographic Quality Assurance Procedures
|
Daily
|
Compare radiographs with reference films.
Enter findings in retake log.
Replenish processing solutions.
Check temperature of solutions.
Make step-wedge test of processing system.
|
|
Weekly
|
Replace processing solution.
Clean processing equipment.
Clean view boxes.
Review retake log (Auditing).
|
|
Monthly
|
Check darkroom safe lighting.
Clean intensifying screens.
Rotate film stock.
Check exposure charts.
|
|
Yearly
|
Calibrate X-ray machine.
|
Key Steps In Radiographic Infection Control
- Apply universal precautions.
- Wear gloves during all radiographic procedures.
- Disinfect and cover X-ray machine, working surfaces, chair and apron.
- Sterilize non–disposable instruments.
- Use barrier protected film (sensor) or disposable container.
- Prevent contamination of processing equipment.
Radiographic Films, Intensifying Screens and Processing
Film Speed
Film speed is controlled largely by the size of the silver halide grains and their
silver content. The silver halide grains are composed mainly of crystals of silver
bromide. Iodide is added to Ultra speed films because of its large diameter (compared
to bromide). It disrupts the regularity of the silver bromide crystal structure,
thereby increasing its sensitivity to X–rays.
The photosensitivity of the silver halide crystals also depends on the presence
of trace amounts of a sulphur containing compound. In addition trace amounts of
gold are sometimes added to silver halide crystals to improve their sensitivity.
Radiographic Speed
It refers to the amount of radiation required to produce an image of standard density.
Film speed frequently is expressed as the reciprocal of the exposure (in roentgens)
required to produce an optical density of 1 above gross fog. A fast film requires
a relatively low exposure to produce a density of 1, whereas a slower film requires
a longer exposure for the processed film to have the same density.
The speed of the dental intra-oral X–ray film is indicated by a letter designating
a particular group. The fastest film available has a rating of ‘F’.
Intensifying Screens
The presence of intensifying screens creates an image receptor system that is 10
to 60 times more sensitive to X–ray than the film alone. The use of intensifying
screens means a substantial reduction in the dose of radiation to which the patient
is exposed. They are used with films in virtually all extra-oral radiography.
|
Emission
|
Phosphor
|
|
Green
|
Terbium activated Gadolinium-oxy-sulphide
|
|
Blue and UV
|
Niobium activated Yttrium tantalate
|
Processing Of X-ray Films
Dark room equipment
The darkroom should be lightproof and well ventilated. There should be optimal conditions
for developing fixing and washing of radiographs.
Safe-lighting
A low intensity illumination of relatively long wavelength (red) that does not rapidly
affect open film but allows adequate workability. To minimize the fogging effect
of prolonged exposure, the safe light should have a 15 watt bulb and should be mounted
at least 4 feet above the surface. The red GBX-2 filter is recommended as a safe
light in dark rooms.
Manual processing tanks
The tank must have hot and cold running water and a means of maintaining the temperature
between 60 and 75 F. The developer is placed on the left side and the fixer is placed
on the right. The master tank should have a cover to reduce oxidation of the processing
solutions, to protect the developing film from accidental exposure to light and
minimize evaporation of the processing solutions. The following time and temperature
combinations have been suggested for optimal developing of X-ray films:
|
Temperature
|
Development Time
|
|
68 F
|
5 minutes
|
|
70 F
|
4 ½ minutes
|
|
72 F
|
4 minutes
|
|
76 F
|
3 minutes
|
|
80 F
|
2 ½ minutes
|
Common Causes and Correction of Unsatisfactory Radiographs
|
Error
|
Cause
|
Correction
|
|
Apical ends of teeth cut off
|
Film is placed too close to teeth in maxillary arch in paralleling technique. Flat
vertical angulation which causes elongation
|
Move film away from the teeth. Increase vertical angulation (in shallow palatal
vault)
|
|
Overlapping of teeth
|
Plane of film not parallel with lingual surface of teeth. Incorrect horizontal angulation
of cone
|
Place film parallel to the teeth and direct central ray of X – ray beam perpendicular
to facial surface of the teeth
|
|
Lesion not showing completely
|
Faulty film placement
|
Center the film over the teeth to be radiographed
|
|
Crowns of the teeth not showing
|
Not enough film showing below or above the crowns of the teeth. Vertical angulation
too steep
|
Increase amount of film showing below and above the crown of the teeth. Decrease
vertical angulation
|
|
Partial image (cone cut)
|
Cone of radiation not covering area of interest
|
Correct horizontal and vertical position of cone
|
|
Shape distortion or fore-shortening
|
Bisecting technique: Vertical angulation of cone too acute. Paralleling technique:
Film not parallel with long axis of teeth. Long cone not positioned correctly
|
Reduce vertical angulation. Place film parallel to long axis of teeth. Position
long cone such that central ray strikes film at right angle
|
|
Elongation
|
Bisecting technique: Vertical angulation of cone too flat. Paralleling technique:
Film not parallel with long axis of teeth. Long cone not positioned correctly
|
Increase vertical angulation. Place film parallel to long axis of teeth. Position
long cone such that central ray strikes film at right angle
|
|
Image distorted
|
Film is bent as patient bites on film holder, bite-block or while patient holds
film in mouth
|
Use a film backing
|
|
Herring bone effect
|
Back side of film placed towards the cone of radiation
|
Place film correctly
|
|
Black dots in apical area
|
Manufacturers identifying mark on film placed towards apical area of teeth
|
Place black dot of film towards occlusal or incisal surface of the teeth
|
|
Artifacts: Writing lines on radiographs, Black marks on radiographs, Black lines
on radiographs
|
Write on film packet with pressure. Moisture contamination. Routine bending of film
to reduce patient discomfort
|
Use less pressure while writing. Blot film packet after removing from patients mouth.
Avoid bending of films
|
|
High contrast
|
Insufficient penetration. Over development. Use of film and / or intensifying screen
of too high contrast. Too long exposure
|
Increase kilovoltage. Use proper technique. Use low contrast or slow speed screens.
Decrease exposure time
|
|
Low contrast
|
Excessive penetration. Under development. Use of film and / or intensifying screen
of low contrast. Scattered radiation
|
Decrease kilovoltage. Use proper technique. Use high contrast or high speed screens.
Check diaphragm size and use appropriate cone
|
|
Fog
|
Light leaks in darkroom. Improper safelight. Radiation: Insufficient protection.
Chemical: High developing temperature. Strong developing solution. Prolonged developing.
Contaminated developer.Deterioration of film
|
Check doors and walls for leaks. Use appropriate safe light. Store unexposed films
in lead protection. Use appropriate developing technique. Store films in good ventilation
and rotate film stock
|
|
Streaks on films
|
Failure to agitate films during development. Chemical deposits on hanger clips.
Excessive drying temperature. Insufficient fixing. Dirty or contaminated water
|
Agitate films when placed in developer solution. Keep clips clean. Reduce air flow
over films. Fix the films properly. Wash films in running water
|
|
Blisters on films
|
Unbalanced processing temperatures. Excessive acidity of fixer. Films not agitated
when immersed in fixer
|
Control processing temperature. Replenish / replace fixer. Agitate films when placed
in fixer
|
|
Reticulation (orange-peel appearance)
|
Extreme temperature changes while processing. Weak fixer solution
|
Maintain uniform processing temperature. Replenish / replace fixer
|
|
Frilling
|
Hot processing solution
|
Maintain uniform processing temperature
|
|
Air bells
|
Air bubbles formed on surface of film
|
Agitate films when placed in fixer
|
|
White spots or lines on films
|
Dirt or dust on film or on screen. Emulsion tears due to rough handling of films
|
Clean intensifying screens periodically and keep darkroom clean. Handle films gently
|
|
Black spots on films
|
Dirt or dust on undeveloped film. Film splashed with water before developing. Films
touching tank during developing
|
Prevent dry chemical powder of developer to contact the film. Careful handling of
films. Films should not touch tank while processing
|
|
Artifacts: Black crescents, Black smudge marks, Black lines
|
Rough handling of films. Finger prints or abrasions. Static electricity
|
Handle film by edges only. Dry fingers before handling films. Remove film from packet
slowly
|
|
Stains on films: Yellow brown stains, Dichroic (two colour)
|
Exhausted developer. Oxidized developer. Prolonged developing. Insufficient rinsing.
Old / exhausted developer. Exhausted fixer. Contamination of developer by fixing
solution. Inadequate fixing. Inadequate rinsing with water
|
Replace developer. Keep developer covered. Use correct time. Rinse film properly.
Replace developer. Replace fixer. Avoid contamination. Fix the film properly. Rinse
film properly
|
|
Deposits on films
|
Contaminated solution. Chemical deposits on hanger. Dirt from dirty water. Metallic
deposits. Fixer contains excessive amount of silver. Milky appearance of fixer due
to aluminum sulfite deposits
|
Change solution.Keep hangers clean. Use running water. Keep tanks covered
|
|
White deposits
|
Fixer contains excessive amount of silver. Milky appearance of fixer due to aluminum
sulfite deposits
|
Change fixer solution. Replenish / replace fixer
|
|
Faded image on radiograph
|
Exhausted fixer. Inadequate fixing. Inadequate final wash
|
Replace fixer. Use adequate time. Wash films adequately
|
|
Brittleness of films
|
Excessive drying temperature. Excessive drying time. Excessive fixer acidity.
|
Reduce dryer temperature. Reduce drying time. Replace fixer solution.
|
|
Overall light films
|
Developer temperature is low. Exhausted developer. Contaminated developer.
|
Use proper developing technique with appropriate time and temperature.
|
|
Overall dark films
|
Developer temperature high
|
Decrease temperature
|
|
Fogged film
|
Developer contaminated by fixer solution. Light leaks. Improper safe light.
|
Change solution. Check for light leaks in darkroom. Use proper safe lights. |
|
Peeling of film or emulsion
|
Developer temperature high.Depleted fixer. Deposits in developer tanks. Improper
film.
|
Decrease temperature. Change fixer. Clean tanks regularly. Check film type and speed
|
|
Pressure marks
|
Improper handling of films
|
Handle films from corners
|
|
Cloudy or smudge appearance on film (greenish / yellow colour)
|
Depleted fixer. Improper type of film
|
Change fixer solution. Check film type
|
|
White cloudy appearance
|
No water in wash tank
|
Change water in tank
|
|
Scratches on film surface
|
Improper handling of films
|
Handle films from corners
|
|
Drying pattern on film surface
|
Dryer too hot. Characteristic of film
|
Reduce dryer temperature. Check type of films
|
Patient Education and The Dental Radiographer
The radiographer must be able to educate patients about the importance of dental
radiographs. The dental radiographer must also be prepared to answer common questions
asked by patients about the need for dental radiographs, X-ray exposure, the safety
of dental X-rays and other miscellaneous concerns.
Legal Issues and The Dental Radiographer
The dentist must be aware of the legal implications involved in taking dental radiographs.
Furthermore, because exposing dental radiographs has implications for patient care,
including the diagnosis of dental disease and treatment planning, the possibility
of negligent care exists when dental radiographs are not properly exposed or used.
Licensure Requirements
The licensed dentist is not required to obtain additional certification to expose
dental radiographs legally.
Informed Consent
Persons seeking health care services, including dental care, have the right to self-determination;
they have the legal right to make choices about the care they receive, including
the opportunity to consent to or to refuse treatment. Therefore, prior to receiving
treatment, the dental patient should be informed of the various aspects of the proposed
treatment, including such diagnostic procedures as the exposure of dental radiographs.
Patient Records
A dental record must be established for every patient and dental radiographs are
an integral part of such a record.
Documentation
It is very important that the dental record include documentation of the exposure
of dental radiographs. The dental record must include documentation of the following:
- Informed consent.
- The number and type of radiographs exposed.
- The
rationale for exposing such radiographs.
- The diagnostic information obtained
from the interpretation of the radiographs.
Confidentiality
All the information contained in the dental record, including dental radiographs,
is confidential or private.
Ownership And Retention Of Dental Radiographs
Legally, dental radiographs are the property of the dentist. The dentist owns the
dental radiographs even though the radiographs were paid for by the patient. Patients
do, however, have a right to reasonable access to their records. When a patient
transfers to another dentist, he or she can request in writing that his or her dental
records be forwarded to that dentist. The original radiographs should not be forwarded;
duplicates should be made and forwarded instead. The patient's written request should
be placed in the dental record as evidence of the patient's directive. Dental records
and dental radiographs should be retained indefinitely.
Patients Who Refuse Dental Radiographs
Patients may refuse dental radiographs. When this occurs, the dentist must carefully
consider the situation. The dentist must then decide whether an accurate diagnosis
can be made and treatment can be provided. In most cases, patient refusal of dental
radiographs compromises the patient's diagnosis and treatment. In such cases, the
dentist cannot treat the patient.
Advances In Imaging Procedures
- Digital Radiography
Digital imaging has revolutionized the field of radio diagnosis. This aims at greater
image enhancement, feature extraction and lesser dose of radiation to patients and
operators and larger gray scale contrast. The chemical processing is eliminated
so there are lesser chances of errors and repeat radiographs. - Orthopantomography
(OPG)
An Orthopantomogram (OPG), also known as an "orthopantogram" or "panorex", is a
panoramic scanning dental X-ray of the upper and lower jaw. It shows a two-dimensional
view of a half-circle from ear to ear. It is commonly used to determine the status
of wisdom teeth. - Computed Tomography
Computed tomography (CT) is a medical imaging method employing tomography. Since
its introduction in the 1970s, CT has become an important tool in medical imaging
to supplement X-rays and medical ultrasonography. Although it is still quite expensive,
it is the gold standard in the diagnosis of a large number of different disease
entities. It is also used for preventive medicine or screening for disease. -
Magnetic Resonance Imaging
Magnetic resonance imaging (MRI), noninvasive diagnostic technique that uses nuclear
magnetic resonance to produce cross-sectional images of organs and other internal
body structures. MRI is used in the diagnosis of brain tumors and disorders, spinal
disorders, multiple sclerosis and cardiovascular disease as it is unhampered by
bone and capable of producing images in a variety of planes. The procedure is considered
to be without risk, but the scanner may interfere with pacemakers, hearing aids
or other mechanical devices. - Ultrasonography
Use of ultrasonic waves to produce images of body structures. The waves travel through
tissues and are reflected back where density differs (e.g. the border between a
hollow organ's wall and its inside). The reflected echoes are received by an electronic
apparatus that measures their intensity level and the position of the tissue reflecting
them. The results can be displayed as still images or as a moving picture of the
inside of the body. Unlike X-rays or other ionizing radiation, ultrasound carries
minimal risk. - Nuclear Medicine Scans
Nuclear scanning uses radioactive substances to see structures and functions inside
your body. Nuclear scans involve a special camera that detects energy coming from
the radioactive substance, called a tracer. Before the test, the patient receives
the tracer, often by an injection. Although tracers are radioactive, the dosage
is small. During most nuclear scanning tests, the patient lies still on a scanning
table while the camera takes images. Most scans take 20 to 45 minutes.
Minimum Qualification and Experience Required for Personnel
|
Radiologist
|
B.D.S. or an M.D.S. degree from a recognized university
|
|
X-ray Technologist
|
X-ray technologists course of minimum one year duration from a recognized institute
|
|
Radiological Safety Officer
|
Certified by the competent authority
|
|
Service Engineer
|
Degree/Diploma in electrical/electronic engineering from a recognized institute
|
Continuing Dental Education
Practitioners should stay informed of new information on radiation safety issues,
as well as developments in equipment, materials and techniques. Dentists and dental
auxiliaries should participate in continuing education in the many aspects of diagnostic
radiology and technology.
International System of Units
Prior to the introduction of International System of Units (abbreviated as SI units)
quantities for radiation measurements were expressed in special units such as roentgen
for radiation exposure and RAD for absorbed dose. Further, the unit REM was exclusively
used for radiation protection dosimetry. The following table provides the conversion
factors for SI / Special Units including their multiples and submultiples used for
routine radiation measurements in X-ray installations.
|
RADIATION QUANTITY
|
S.I. UNIT (AND SYMBOL)
|
SPECIAL UNIT
|
|
Absorbed Dose
|
1 gray (Gy)
1 centigray (cGy)
|
100 rad
1 rad
|
|
Dose Equivalent
|
1 Sievert (Sv)
10 millisievert (mSv)
10 microsievert (Sv)
|
100 rem
1 rem
1 millirem
|
Certain measurements, which were performed in exposure quantity and expressed in
its special unit the roentgen, have now been replaced by air kerma measurements
in free air. An exposure of 1 roentgen is equivalent to an air kerma of 8.7 milligray
(mGy). An air kerma of 10 microgray per hour (Gyh-1) corresponds to an exposure
rate of 1.15 milliroentgen per hour (mR/hr).
S.I. PREFIXES
Mirco () = 10-6 Mega (M) = 106
Milli (m) = 10-3 Giga (G) = 109
Centi (c) = 10-2 Tera (T) = 1012
Glossary of Terms
Adequate Protection means protection against radiation to keep levels of
radiation, As Low As Reasonably Achievable (ALARA Principle) and in no case exceed
the prescribed operational limits.
Dose means energy absorbed in matter from ionizing radiation per unit mass
of the matter. The SI unit of dose is Gray (Gy). Special unit of dose is rad. 1
rad = 1 cGy
Dose Equivalent means the quantity obtained on multiplying the absorbed dose
in tissue by appropriate weighting factors to correct and normalize for variation
in the degree of biological effect produced by the same dose of different ionizing
radiations or under different irradiation conditions. The unit is Sievert (Sv) :
1 Sv = 100 rem
Operational Limits means limits on levels of radiation as the competent authority
may specify by notification from time to time.
Quality Assurance Tests means tests performed to ensure the performance and
reliability of X-rays equipment as per the design specifications.
Protective Barrier or Shielding means a barrier of radiation attenuating
material used to reduce radiation levels.