Infection Control During Clinical Procedure
Preparation Of Operator
The process of infection control should begin before the patient arrives, during
preparation for clinical treatment. This reduces the risk for transmission of infectious
agents during patient care, makes the treatment session more efficient and the post
treatment infection control process easier and more effective.
Infection Control During The Pretreatment Period (For Operatory)
- Remove unnecessary items and unused equipments from the operatory.
- The operatory should be arranged to facilitate thorough cleaning following each
patient.
- Use prearranged tray set-ups for routine or frequently performed procedures whenever
possible.
- Preplan the materials needed during treatment. Set out all instruments, medications
and impression materials needed. This includes individually sterilized bur blocks
with only those burs needed for the procedure. Also put out a rubber dam setup if
one will be needed.
- Use disposable items and unit-dose materials whenever possible. The use of disposable
items saves time during cleanup and decontamination. Disposables also solve the
problem of decontaminating hard-to-clean items such as the saliva ejector.
- Identify equipment and surfaces that will become contaminated during treatment.
Use a barrier. to prevent contamination of these surfaces and items or to disinfect
them after treatment.
- The decision to use barriers or chemical disinfection should be based on individual
circumstances. Barriers are quick and easy to use, and can be readily changed, but
may be more expensive than chemical disinfection.
- Disinfectants are generally less expensive and are easy to use on flat surfaces,
but they may stain or corrode some materials, may be toxic and are difficult to
use effectively on rough and odd-shaped surfaces.
- If barriers are chosen, a number of readily available materials can be used. These
include plastic wraps, aluminum foil, impervious backed paper and commercially available
polyethylene sheets and tubing.
Some examples of the use of barriers are as follow
- Cover light handles with plastic wrap or aluminum foil. These are available commercially
and some manufacturers offer removable light handle attachments that can be disinfected
easily.
- Cover the back of chair with a polyethylene bag to protect the headrest, protect
the control buttons with plastic wrap and the arms of the chair with polyethylene
tubing.
- Cover counter tops with plastic laminated paper.
- Protect the hoses to the air-water syringe, saliva ejector, high- speed evacuator
and handpieces with polyethylene tubing.
- Set up radiographs for viewing and reviewing patient records before initiating the
treatment. Entries into the record should be made before putting the gloves on or
after they have been removed and hands have been washed. Do not make computer entries
during treatment unless the equipment has been barrier protected.
- Follow manufacturer's directions for care and maintenance of water lines. Patient
material (e.g. oral microorganisms, blood, saliva) can enter the dental water system
during patient treatment. Dental devices connected to the dental unit water system
should be operated to discharge water and air for a minimum of 20-30 seconds between
patients. This procedure flushes out any patient material that might have entered
the turbine, air and waterlines.
- All personnel involved in patient care should prepare themselves for the incoming
patient. This includes the use of personal protective equipment (gown, eyewear,
mask, gloves) and hand washing.
Infection Control During Clinical Procedures
Infection Control practices for use during clinical activities will help reduce
the exposure to infectious agents.
Infection Control Practices-Chair side
- Use care when receiving, handling or passing sharp instruments. Many dental instruments
can easily cut gloves and skin.
- Take special precautions with syringes and needles. Needle stick injuries are a
major cause of infection in health care personnel. Needles should not be recapped,
bent, broken or otherwise manipulated by hand.
- Never recap a needle using a two-handed technique. Instead, use one of the commercially
available sheath holders or the "scoop" technique. In this technique, the cap is
scooped up from the tray with the needle tip using only one hand. As an additional
protection against needle sticks, do not allow uncovered needles to remain on the
instrument tray. It is far safer to dispose of them immediately after use in a puncture-
proof container
- Disposable needles should not be bent or broken after use.
- Needles should not be removed manually from disposable syringes or otherwise handled
manually.
- Forceps or other appropriate instruments may be used to handle sharp items.
- Disposable syringes, needles, scalpel blades and other sharp items should be discarded
into puncture-resistant biohazard (sharps) containers that are easily accessible.
- Use a rubber dam whenever possible. Rubber dams limit the splash and spatter of
blood and saliva and should be used whenever possible .
- Avoid touching unprotected switches, handles and other equipment after gloves have
become contaminated. If objects are touched or handled, they should be carefully
cleaned and disinfected at the end of the procedure.
- Avoid opening drawers or cabinets once gloves have become contaminated.
- If it becomes necessary, you may simply ask another person for assistance; or you
may use another barrier, such as prepackaged aluminum foil squares or plastic gloves,
to grasp the cabinet or drawer handle.
- However, if these options are not available, you must remove the contaminated gloves,
wash hands before opening a drawer or cabinet, and then reglove before resuming
patient treatment.
- Extracted teeth should be considered as potentially infectious material and should
be disposed in medical waste containers.
Pre-procedural Mouth Rinses
- Anti-microbial mouth rinses (e.g. chlorhexidine gluconate, essential oils, or povidone-
iodine) used by patients before a dental procedure are intended to reduce the number
of microorganisms the patient might release in the form of aerosols or spatter that
subsequently can contaminate dental practitioner and equipment operatory surfaces.
Preparation Of Operator Between Patient/Clinical Session
The infection control process continues after the patient leaves the office. Although
effective pretreatment planning will simplify your task, there are a number of tasks
that should be done following patient care to further reduce the risk for transmission
of infectious agents.
Infection Control Procedures-Post treatment
- Continue to wear personal protective equipment during cleanup. Begin the cleaning
and disinfection process by removing contaminated gloves used during treatment.
- Next, wash your hands and put on a pair of utility gloves before beginning clean-up.
Continue to wear protective eyewear, mask and gown.
- Remove all disposable barriers. All barriers placed before treatment, including
light handle covers, polyethylene tubing and counter-top barriers, should be removed.
These should be placed into a leak-proof waste bag inside a trash container.
- Dispose of blood and suctioned fluids that have accumulated in the collection bottles
during treatment.
- Identify a utility sink, drain or toilet connected to a sanitary sewer that can
be used solely for the disposal of blood, liquid wastes and suctioned fluids.
- After the blood and suctioned fluids are poured into the sink, use a 1 : 100 dilution
of household bleach (or other appropriate intermediate-level disinfectant) and tap
water to disinfect the dental unit collection bottle.
- The bottle should be completely filled and the solution should be kept in the bottle
for 10 minutes (or according to manufacturer's instructions) before emptying and
rinsing with fresh water.
- Clean and disinfect all items not protected by barriers. All clinical contact surfaces
that were not protected by barriers must be cleaned and then disinfected with an
intermediate-level surface disinfectant- for example iodophor, phenolic solution,
or diluted household bleach (sodium hypochlorite).
- Remove personal protective equipment. The proper method of removing a mask is to
grasp it only by the cloth or elastic strings, not by the mask itself.
- Reusable protective eyewear and face shields should be cleaned with soap and water
as necessary and, if visibly soiled, disinfected between patients according to manufacturer's
instructions.
- Remember not to touch the eyewear or face shield with ungloved hands, because it
may have become contaminated with spatter of blood and saliva during patient care.
- Protective clothing should be changed when it becomes visibly soiled and as soon
as feasible if penetrated by blood or OPIM.
- Soiled protective gowns should be placed in a soiled linen container.
- Gowns may be washed using the normal laundry cycle.
- Utility gloves should be washed with soap before removal.
- Finally, thoroughly wash your hands for at least 15 seconds, using soap or anti
microbial hand wash.
- Clean the instruments and prepare them for sterilization. Remove the tray with all
instruments to the receiving and clean-up section of the central processing area.
- Instruments should be picked up individually. Never pick up a handful of instruments
because this greatly increases the risk of cuts or punctures.
- Special care should be used when handling double-ended instruments. Don't reach
into trays or containers holding sharp instruments that are not clearly visible.
Use strainer- type baskets to remove them.
- Dental instruments should be scrubbed carefully using a long- handled brush with
soap and water and rinsed thoroughly. As an alternative, instruments may be cleaned
in an ultrasonic cleaner or washer- disinfector. After drying, the clean instruments
are ready for sterilization.
- Sterilize devices connected to the air or waterlines of the dental unit between
each patient. Any device connected to the dental unit air or water system that enters
the patient's mouth (e.g., handpieces, prophylaxis angles, air abrasion devices)
should be run to discharge water and/or air for 20-30 seconds after each patient.
- Follow the manufacturer's recommendations for proper flushing of devices and for
the use and maintenance of waterlines and check valves as appropriate.
- An ultrasonic cleaner should be used to remove any adherent material, but only if
recommended by the device's manufacturer.
- Otherwise, scrub devices thoroughly with a detergent and hot water and dry. Finally,
sterilize it according to the manufacturer's instructions.
- Disposal of any general medical waste (e.g. gloves, masks, lightly soiled cotton
balls, protective barriers) classified as non-regulated medical waste and can be
disposed of along with ordinary waste.
- A single leak-resistant biohazard bag is usually adequate for non-sharp regulated
medical waste. Sharp items, such as needles and scalpel blades, should be placed
intact into puncture- resistant biohazard containers and disposed.
- Disposal of regulated medical waste. (e.g. solid waste soaked or saturated with
blood and saliva, surgically removed hard and soft tissues, and contaminated sharp
items). A single leak- resistant biohazard bag is usually adequate for non-sharp
regulated medical waste. Sharp items, such as needles and scalpel blades, should
be placed intact into puncture- resistant biohazard containers and disposed.
Appropriate procedures for handling sharp instruments include the following
- Wear sturdy utility gloves when cleaning contaminated instruments or other sharp
items.
- Dispose of needles and other sharp items promptly and appropriately.
- When handling sharp instruments, avoid any quick motions that would bring one hand
toward the other or the instrument across any part of your body.
Infection Control During Radiographic Procedures
Dental Radiology
- When taking radiographs, the potential to cross-contaminate equipment and environmental
surfaces with blood or saliva is high if aseptic technique is not practiced.
- Gloves should be worn when taking radiographs and handling contaminated film packets.
- Other PPE (e.g. mask, protective eyewear and gowns) should be used if spattering
of blood or other body fluids is likely to occur.
- Heat-tolerant versions of intraoral radiograph accessories are available and these
semicritical items (e.g. film-holding and positioning devices) should be heat-sterilized
before patient use.
- After exposure of the radiograph and before glove removal, the film should be dried
with disposable gauze or a paper towel to remove blood or excess saliva and placed
in a container (e.g. disposable cup) for transport to the developing area.
- However, care should be taken to avoid contamination of the developing equipment.
- Protective barriers should be used, or any surfaces that become contaminated should
be cleaned and disinfected with a disinfectant (intermediate-level).
- Radiography equipment (e.g. radiograph tubehead and control panel) should be protected
with surface barriers that are changed after each patient.
- If barriers are not used, equipment that has come into contact with dental practitioners
gloved hands or contaminated film packets should be cleaned and then disinfected
after each patient use.
For Digital Radiography
- Digital radiography sensors and other high-technology instruments (e.g. intraoral
camera, electronic periodontal probe, occlusal analyzers and lasers) come into contact
with mucous membranes and are considered semicritical devices.
- They should be cleaned and ideally heat-sterilized or high- level disinfected between
patients.
- Semicritical items that cannot be reprocessed by heat sterilization or high- level
disinfection should, at a minimum, be barrier protected to reduce gross contamination
during use.
- To minimize the potential for device-associated infections, after removing the barrier,
the device should be cleaned and disinfected with a disinfectant (intermediate-level)
after each patient.
- Manufacturers should be consulted regarding appropriate barrier and disinfection/sterilization
procedures for digital radiography sensors, other high-technology intraoral devices
and computer components.
Checklist
Steps to be followed:
- Keep operatory free of unnecessary items.
- Set up all supplies and instruments for a procedure in advance.
- Use disposable items whenever appropriate.
- Use individualized, sterilized bur blocks.
- Set up radiographs and review patient records before treatment.
- Flush waterlines between patients. Wear gown, protective eyewear, masks and gloves
during procedures involving spatter.
- Use proper handwashing techniques.
- Use rubber dams.
- Use high-volume evacuation.
- Use barriers to protect equipment (including digital radiographic equipment) from
contamination whenever possible.
- Keep radiographs and records protected during treatment.
- Use a safe method for recapping needles.
- Disinfect all unprotected clinical contact surfaces in dental operatory.
- Use appropriate methods of handling and passing sharp items.
- Use safe procedures for exposing and developing radiographs.
- Wear protective gear during post treatment clean-up.
- Proper disposal of regulated medical waste.
|
|